Kev ntsuas tsis paub meej ntawm kev ntsuas qhov tshwm sim ntawm qhov ntsuas qhov yuam kev ntawm Lub Teeb Meter

Apr 15, 2023

Kev ntsuas tsis paub meej ntawm kev ntsuas qhov tshwm sim ntawm qhov ntsuas qhov yuam kev ntawm Lub Teeb Meter

 

Photometers tau siv dav hauv kev ntsuas teeb pom kev zoo hauv kev kho mob hauv nroog, kev lag luam thiab kev ua liaj ua teb, thiab kev tsim vaj tsev. Raws li cov kev cai ntawm JJG245-2005 "Illuminance Meter Verification Regulations", lub teeb calibration txoj kev yog saws, thiab lub illuminance uas tsim los ntawm lub teeb siv standard teeb nyob rau hauv ib tug tej yam deb yog siv los calibrate lub 'meter' nyob rau hauv kev sim. Cov khoom siv hluav taws xob hluav taws xob muaj xws li lub teeb pom kev siv hluav taws xob, lub teeb pom kev nrog lub ntsuas (xws li lub ntsuas hluav taws xob, lub diaphragm, lub teeb tuav, thiab lub dav hlau filament hloov kho), thiab DC tswj fais fab mov (cov duab digital). Tag nrho cov txheej txheem calibration yuav tsum tau nqa tawm hauv chav tsaus.


1 Kev tshuaj xyuas cov ntsiab lus tseem ceeb cuam tshuam rau kev ntsuas ntsuas
Raws li cov kauj ruam tseem ceeb uas tau teev tseg hauv cov cai pov thawj, sib sau ua ke 2856K thawj-chav kawm qauv teeb ntawm luminous siv thiab lub illuminance meter nyob rau hauv kev tshuaj xyuas ntawm qhov chaw ntawm lub photometric ntsuas ntsuas. Teeb lub teeb txheem thiab preheat nws. Tom qab qhov tam sim no ntawm tus qauv teeb yog ruaj khov, hloov qhov kev ncua deb ntawm tus qauv teeb thiab lub photometric lub taub hau, nyeem cov zaub tus nqi ntawm lub illuminance meter, xam qhov yuam kev ntawm cov zaub tus nqi, thiab ntsuam xyuas nws tsis meej. Lub teeb meter calibration muaj nruj heev rau cov neeg ua haujlwm thiab cov khoom siv. Cov txheej txheem ntawm tus neeg teb xov tooj thiab kev xaiv cov khoom siv yuav cuam tshuam rau kev ruaj ntseg ntawm lub teeb ci hauv cov qauv txheej txheem, kev tiv thaiv kev cuam tshuam ntawm lub teeb tsis pom kev, kev ntsuas hluav taws xob, ntsuas qhov deb, thiab qhov sib txuam ntawm lub teeb meter lim.


1.1 Ntsuas qhov deb thiab lub teeb khiav
Tus nqi hloov pauv ntawm lub teeb pom kev hauv chav kuaj pom feem ntau ua tiav los ntawm txoj kev teeb pom kev calibration.
Txoj kev no ua rau kev siv tag nrho ntawm txoj cai rov qab square ntawm kev ncua deb: E= 1/2, yog li qhov linearity ntawm kev ntsuas qhov deb thiab lub teeb khiav yuav qhia qhov yuam kev.
Tag nrho qhov yuam kev hauv 1m ntawm qhov ntsuas qhov sib txawv yuav tsum tsis pub tshaj 0.2mm, thiab qhov yuam kev linearity ntawm lub teeb nws tus kheej yuav tsum tsis pub tshaj ± 1mm.
1.2 Filament dav hlau thiab optical taub hau txais qhov chaw
Calibration xav kom cov neeg ua haujlwm txawj thiab paub txog kev kho qhov muag, kev sib tw thiab lwm yam haujlwm. Siv cov cuab yeej sib dhos los kho qhov tau txais qhov chaw ntawm lub teeb pom kev zoo, lub dav hlau ntawm cov qauv teeb pom kev zoo, thiab lub aperture ntawm lub diaphragm kom lawv perpendicular mus rau optical axis, thiab txhua qhov chaw nyob ntawm qhov chaw kho qhov muag. Ntawm qhov tod tes, ua kom tiav txoj kev kho qhov muag kom txo tau qhov kev ua yuam kev los ntawm kev hloov kho tsis raug ntawm lub dav hlau filament thiab qhov tau txais qhov chaw ntawm lub taub hau kho qhov muag.


1.3 Standard teeb qhov chaw
Vim tias lub teeb pom kev ntawm lub teeb siv hluav taws xob siv hluav taws xob siv rau kev ntsuas ntsuas, qhov tseeb ntawm lub teeb pom kev zoo ntawm cov qauv teeb yuav tsum nruj me ntsis txwv. Qib {{0}} luminous siv cov qauv teeb, uas nws kab rov tav thiab ntsug lub kaum sab xis yuav tsum ua tau raws li lub kaum sab xis, cov qauv teeb tig ± 1.5 degree hauv kab rov tav lossis ± 1.0 degree hauv txoj kev ntsug , thiab qhov hloov pauv amplitude ntawm luminous siv yuav tsum tsis pub tshaj 0.6 feem pua ​​. Kev hloov pauv txhua xyoo Tsis pub tshaj 0.7 feem pua.


1.4 Kev ntsuas hluav taws xob
Xa mus rau cov kev cai lij choj kom debug lub DC stabilized fais fab mov: cov zis voltage yog tsis tu ncua adjustable, thiab amplitude ntawm cov zis voltage hloov nyob rau hauv kaum feeb yuav tsum tsis pub tshaj 0.02 feem pua ​​; tom qab tus qauv teeb yog ignited, nws yuav tsum tau sov thiab stabilized ua ntej ntsuas kom paub meej lub luminescence ntawm tus qauv teeb lub zog yog zoo reproduced.
Txawm hais tias tus nqi tam sim no ua haujlwm ntawm tus qauv teeb pom kev zoo raws li tus nqi tam sim no thaum kuaj xyuas, cov khoom siv ntsuas hluav taws xob zoo ib yam feem ntau tsis siv, yog li qhov tseeb ua haujlwm tam sim no tus nqi ntawm cov qauv teeb feem ntau sib txawv, uas yuav qhia txog qhov yuam kev.


1.5 Spectral yam ntxwv ntawm lub teeb meter lim
Thaum ntsuas illuminance, lub spectral responsivity faib s(λ) ntawm lub detector yuav tsum tau raws li lub photopic efficiency v(λ) ntawm lub International Commission on Illumination (CIE), thiab silicon photodiode los yog silicon photocell siv nyob rau hauv lub illuminance meter yog zoo ib yam. nrog v (λ) tsis sib haum, thiab yuav tsum tau ntxiv cov lim dej kho kom raug. Txawm li cas los xij, qhov sib tw s(λ) thiab v
(λ) tsis tuaj yeem ua tau zoo ib yam, thiab kev sib txuam tsis tiav yuav ua rau cov txiaj ntsig tsis raug.


1.6 Stray light shielding
Cov ntaub ntawv ntawm cov cuab yeej txhawb nqa yuav tsum muaj peev xwm tiv thaiv lub teeb pom kev zoo, xws li txav apertures, curtain apertures, thiab lwm yam, kom tiv thaiv lub teeb tsis pom kev los ntawm kev sib xyaw rau hauv txoj kev kho qhov muag thiab rhuav tshem cov txiaj ntsig kev sim. Hauv kev sim tiag tiag, kev tiv thaiv ntawm lub teeb pom kev sab nraud yog nyob rau hauv qhov chaw, tab sis lub teeb tsis pom kev uas tsim los ntawm cov khoom pov thawj nws tus kheej tsis tuaj yeem tiv thaiv tag nrho, uas cuam tshuam nrog kev nyeem ntawv ntawm lub ntsuas hluav taws xob.

 

-3

Xa kev nug