Tej yam yuav tsum nco ntsoov ua ntej siv thundermagnetic pH meter:
Screw lub electrode qia mus rau hauv lub qhov (socket) electrode, kho lub electrode clamp mus rau txoj hauj lwm tsim nyog, thiab tshem tawm lub qhov hluav taws xob luv luv.
Cov khoom siv hluav taws xob sib txuas yog rub tawm ntawm lub electrode clip thiab lub tes tsho electrode nyob rau pem hauv ntej kawg ntawm lub electrode yog rub tawm. Yaug lub electrode nrog dej distilled thiab blot qhuav nrog lim ntawv tom qab ntxuav. Txuas lub zog rau hauv lub qhov hluav taws xob.
Tom qab lub hwj chim qhib, preheat rau 30 feeb, thiab ces ua calibration. Cov cuab yeej yuav tsum tau calibrated ua ntej siv. Feem ntau, thaum lub cuab yeej siv tas li, nws yuav tsum tau ntsuas ib zaug ib hnub.
Thunder magnet pH meter calibration kauj ruam
Txuas cov hluav taws xob sib xyaw rau hauv qhov ntsuas hluav taws xob ntsuas;
Kho tus xaiv hloov pob qhov rooj mus rau PH txoj hauj lwm;
Nias lub pob "Tub kub" los muab cov cuab yeej tso rau hauv qhov kev daws teeb meem kub hloov lub xeev. Nias "▲" lossis "▼" ntawm qhov "Tsub" tus yuam sij los kho qhov ntsuas kub kom nce lossis poob, kom qhov ntsuas kub ntawm qhov ntsuas yog qhov ntsuas kub tam sim no, thiab tom qab ntawd nias "Qhib" "qhov tseem ceeb, ntsuas rov qab mus rau qhov ntsuas pH tom qab txiav txim siab qhov ntsuas kub.
Siv lub ntsuas pH sib nqus los ntxig cov electrode uas tau ntxuav nrog dej distilled thiab qhuav nrog cov ntawv lim rau hauv cov qauv kev daws teeb meem ntawm pH=6.86. Tom qab kev nyeem ntawv ruaj khov, nias lub "txoj hauj lwm" qhov tseem ceeb (pH qhov taw qhia lub teeb nyem maj mam nyob rau lub sijhawm no, qhia tias lub ntsuas ntsuas nyob rau hauv qhov chaw thiab calibration lub xeev), nias "▲" lossis "▼" ntawm qhov "Qhov chaw" qhov tseem ceeb los kho cov pH zaub tus nqi kom nce lossis poob, kom lub ntsuas ntsuas kev nyeem ntawv zoo ib yam nrog tus nqi pH ntawm cov tshuaj tsis haum rau lub sijhawm ntawd, thiab tom qab ntawd nias lub pob "Confirm" (xws li cov phosphoric acid sib xyaw thaum ntsev yog 10 degree. , pH =6.92).
Tig txoj kab nqes kev kho qhov rooj clockwise mus rau hauv qab (uas yog, mus rau 100% txoj hauj lwm); kho qhov chaw hloov kho pob qhov rooj kom lub ntsuas ntsuas kev nyeem ntawv zoo ib yam nrog tus nqi pH ntawm cov tshuaj tsis haum thaum qhov kub thiab txias tsis tu ncua (piv txwv li, thaum siv cov phosphoric acid sib xyaw los ua qhov ntsuas kub ntawm 100C, PH=6.92 )
Ntxig cov electrode uas tau ntxuav nrog dej distilled thiab qhuav nrog cov ntawv lim rau hauv cov qauv kev daws teeb meem ntawm pH=4.00 (los yog pH=9.18). Tom qab kev nyeem ntawv ruaj khov, nias lub pob "Slope" thiab nias "▲" lossis "▼" ntawm tus yuam sij "Slope". Kho qhov pH zaub tus nqi kom nce lossis poob kom qhov kev nyeem ntawv yog pH tus nqi ntawm cov tshuaj nyob rau lub sijhawm ntawd, thiab tom qab ntawd nias lub pob "Qhib". Lub ntsuas pH nkag mus rau hauv lub xeev ntsuas pH, pH qhov taw qhia lub teeb nres flashing, thiab lub ntsuas ntsuas ua tiav calibration. Kev ntsuas tsuas tuaj yeem ua tau tom qab ua tiav calibration.