Cov ntsiab lus ntawm cov laj thawj rau Digital Instability ntawm PH Meter
1. Xyuas seb cov electrode puas lawm;
2. Nws yuav tsum yog tias cov electrode tau siv ntev dhau lawm, ua ntej calibrate kom pom tias nws zoo;
3. Koj tuaj yeem sim so qhov kev sojntsuam nrog 2.5MMOL / L KCL tov;
4. Ntxuav lub khob iav, puas yog lub sij hawm ntev, qee cov organic teeb meem txuas rau nws, ua rau cov tshuaj tiv thaiv tsis zoo;
5. Muaj cov tshuaj sib npaug hauv dej ~ CO2 ntxiv rau H2O → H ntxiv rau HCO3-, txij li cov dej ntshiab huv lossis cov dej saum npoo av tsis muaj zog alkaline, qhov sib npaug yuav txav mus rau qhov kev taw qhia zoo, yog li pH yuav ib txwm sawv ~ Kuv tus kheej xav tias nws yog Xws li;
6. Ntxiv cov ntsev nruab nrab (xws li Kcl) ua lub zog ionic regulator rau cov qauv dej kuaj kom hloov tag nrho cov ionic zog hauv cov tshuaj, ua kom cov conductivity, thiab ua kom qhov ntsuas ceev thiab ruaj khov. Txoj kev no tau teev tseg nyob rau hauv lub teb chaws tus qauv GB/T6P04.3-93: "Yuav kom txo cov kev cuam tshuam ntawm cov kua hlws ris muaj peev xwm thiab ua kom muaj kev ruaj ntseg sai thaum ntsuas cov qauv dej, ntxiv ib qho poob nruab nrab {{ 7}}.1moL/LKCL tov rau txhua 50ml dej qauv. Txawm hais tias txoj kev no Hloov lub zog ionic hauv cov qauv dej ua rau nws tus nqi pH hloov mus rau qee qhov, tab sis kev sim tau ua pov thawj tias qhov kev hloov pauv no tsuas yog hloov pauv ntawm 0.01pH hauv tus nqi, uas yog qhov ua tau tiav. Txawm li cas los xij, thaum siv txoj kev no, nws yuav tsum raug sau tseg tias cov tshuaj Kcl ntxiv yuav tsum tsis txhob muaj cov alkaline lossis acidic impurities. Yog li ntawd, Kcl reagent yuav tsum muaj siab purity, thiab cov dej zoo ntawm cov tshuaj yuav tsum yog high-purity nruab nrab dej zoo;
7. Thaum lub sij hawm ntsuas, CO2 yog absorbed, thiab pH nce ntxiv;
8. pH ntsuas pH tus nqi, lub hauv paus ntsiab lus yog tias ib qho hluav taws xob tsim los ntawm cov khoom siv hluav taws xob thiab cov khoom siv hluav taws xob tau muab tso rau hauv cov tshuaj los tsim cov roj teeb thawj. Nyob rau hauv chav tsev kub (25 degree), txhua chav tsev ntawm pH tus nqi yog sib npaug rau electromotive quab yuam hloov tus nqi ntawm 59.1mv. Qhov ntsuas tau qhia ncaj qha raws li kev nyeem pH, thiab qhov sib txawv ntawm qhov kub thiab txias yog muab nrog cov khoom siv them nyiaj ntawm lub ntsuas. Vim tias cov dej ntshiab muaj cov ions tsawg heev thiab tsis tuaj yeem tsim lub roj teeb ruaj khov, cov ntsev nruab nrab (xws li KCl) tau ntxiv rau cov qauv dej kom raug kuaj ua tus tswj lub zog ionic los hloov tag nrho cov ionic zog hauv cov tshuaj, nce cov conductivity. , thiab ua kom qhov ntsuas ceev thiab ruaj khov;
9. Kev nyeem ntawv ntawm ph meter tsis ruaj khov: qhov ntsuas ntsuas yog acidic, thiab txoj kab nqes yog kho nrog cov tshuaj tsis haum pH =4, thiab cov tshuaj ntsuas yog alkaline, thiab txoj kab nqes yog kho nrog cov tshuaj tsis haum. ntawm pH =9. Qhov ze ntawm tus nqi pH ntawm cov tshuaj rau kev kho qhov nqes hav yog mus rau pH tus nqi ntawm cov tshuaj ntsuas, qhov zoo dua;
10. Tej zaum yuav muaj kev sib cuag tsis zoo;
11. Yog hais tias pH meter co me ntsis, kev nyeem ntawv yuav hloov;
12. Nws yuav tsum tsis txhob yog vim li cas rau electrode. Kuv xav tias koj cov electrode qhuav. Koj tuaj yeem pom ntawm koj tus kheej. Feem ntau, muaj ib lub qhov me me ntawm lub electrode. Muab qee qhov 3mol / LKCL rau hauv nws thiab nws yuav tsum ua tiav;
13. Tus nqi pH tsis ruaj khov ntawm cov electrode thaum siv qhov pib tsis muaj dab tsi cuam tshuam nrog lub calibration. Nws muaj feem xyuam rau qhov hloov pauv ntawm lub zog hluav taws xob, kev ua haujlwm ntawm lub tshuab hluav taws xob, cov hlau lead ntawm cov electrode, kev sib cuag ntawm lub tshuab hluav taws xob, thiab qhov ntsuas kub ntawm cov tshuaj yuav tsum ntsuas. Thaum calibrating, yog tias qhov ntsuas ntsuas ze rau acidic, siv "6" rau qhov chaw, thiab yog tias nws yog alkaline, siv "9" rau qhov chaw. Ob leeg tsis tuaj yeem xaiv los ntawm kev txiav txim siab;
14. Nws yuav tsum muaj teeb meem nrog cov electrode, qhib nws ua ntej siv. Ntxiv mus, tsis hais seb cov electrodes siv los yog tsis, lawv yuav raug tshem tawm hauv ib xyoos. Rau cov kua qaub, siv buffers ze rau 4 thiab 7 rau kev kho; rau alkaline daws, siv buffers 7 thiab 10 rau kev kho;
15. Nws yog qhov txawv ntawm pH ntawm cov dej ntshiab tsuas yog ntsuas los ntawm pH meter. Ua ntej tshaj plaws, nws yog ib qho tsim nyog yuav tsum nkag siab txog lub hauv paus ntsiab lus ntawm kev ntsuas pH meter: cov ions uas muaj nyob rau hauv cov dej txav mus rau hauv qhov kev txiav txim ntawm electrode los ua ib qho tam sim no los tso saib nyeem ntawv. Cov dej ntshiab dhau ob peb ions, tsawg electrolysis tus nqi, qhov ntsuas yuav ruaj khov li cas;
16. Tej zaum nws yuav yog kev sib cuag tsis zoo. Los yog cov electrode soaking tov (3MOL / L KCL tov) tsawg dua, thiab cov electrode tsis yog tag rau hauv cov electrode soaking tov. Tsis tas li ntawd, cov electrode yog laus, los yog nws yuav tsum tau hloov;
17. Kuv xav tias nws muaj ib yam dab tsi ua rau chav tsev kub. Thaum qhov kub thiab txias qis los yog huab cua khiav ceev, nws yuav cuam tshuam rau pH meter. Chav tsev kub yuav tsum ruaj khov, thiab lub qhov rooj thiab qhov rais yuav tsum kaw thaum ntsuas;