+86-18822802390

Karl Fischer Moisture Analyzer Solution rau Over-titration Endpoint

May 11, 2023

Karl Fischer Moisture Analyzer Solution rau "Over-titration Endpoint"

 

Thaum Karl Fischer noo noo analyzer (tseem hu ua: insulating roj micro-dej ntsuas ntsuas, noo noo analyzer) muaj titration tsis ua hauj lwm, ntxiv rau tsis tuaj yeem ncav cuag lub xeev sib npaug thiab ua qhov kev soj ntsuam ib txwm, xim ntawm Karl Fischer reagent hauv lub titration tank kuj yuav hloov. Pom tsis pom kev tshwm sim (cov ntsiab lus sib npaug yog tsaus dua).


Feem ntau ua rau overtitration tsis ua hauj lwm nyob rau hauv Karl Fischer ya raws analyzers


1. Tom qab siv lub twj tso kua mis, tsuas yog lub zog ntawm lub stirrer yog muab tua, tab sis lub ntsuas ntsuas tsis tau muab tua, yog li ntawd lub generating electrode tseem nyob rau hauv kev ua hauj lwm, thiab ntau tshaj iodine yog electrolyzed los ntawm lub reagent, uas tshaj lub equilibrium lub xeev ntawm qhov ntsuas.


2. Lub nplawm ceev yog tsawg dhau, thiab cov khoom siv iodine uas ua los ntawm electrolysis tsis sib xyaw rau lub sijhawm. (Cov tuam txhab sib txawv thiab cov qauv ntawm reagent formulations, sib txawv reagent viscosity, ceev)


daws teeb meem
Tom qab pom muaj teeb meem, tsis txhob nchuav Karl Fischer reagent hauv lub ntsuas thiab hloov tam sim ntawd.


1. Tig lub zog ntawm lub stirrer thiab cia cov iodine ntau dhau do tusyees nyob rau hauv lub pas dej titration. Ua kom cov kev sib hloov ceev (xaiv qhov tsim nyog stirring ceev, qhov ceev yuav cuam tshuam rau kev siv cov reagents)


2. Siv lub micro-syringe los rho cov dej ultrapure, thiab tom qab ntawd ntxiv qhov tsim nyog ntawm cov dej ultrapure mus rau qhov ntsuas, cov xim ntawm cov reagent yuav maj mam sib zog, kom txog thaum nws dhau lub xeev dej. Pib electrolysis kom rebalance thiab mus nrog kev tsom xam li qub.

 

Moisture Meter

 

 

 

 

Xa kev nug