+86-18822802390

Kev ntsuas pH meter malfunction piv txwv

Aug 07, 2024

Kev ntsuas pH meter malfunction piv txwv

 

1. Thaum lub sij hawm siv lub ntsuas acidity, nws tau pom tias qhov ntsuas tsis tuaj yeem raug xoom thaum kho qhov xoom potentiometer, tab sis tus lej tuaj yeem hloov pauv.
Kev tshuaj xyuas thiab kev soj ntsuam yuam kev: Raws li ib txwm muaj, lub twj paj nruag tuaj yeem dhia los ntawm kev kho lub xoom potentiometer. Xyuas lub preamplifier thiab siv ib tug multimeter los ntsuas DC voltage ntau ntawm lub preamplifier integrated block. Nws tau pom tias qhov hluav taws xob tsis tuaj yeem hloov kho, qhia tias muaj kev cuam tshuam hauv Circuit Court hauv cheeb tsam no. Kev ntsuas ntxiv tau qhia tias nws yog vim qhov tsis sib haum xeeb ntawm cov hluav taws xob hluav taws xob hauv hluav taws xob. Hloov qhov tsis kam ntawm qhov sib txuam ua haujlwm-tsim transistors kom phim.


Kev daws teeb meem: Tom qab hloov, lub ntsuas tau muab tso rau hauv ib txwm siv.


2. Lub ntsuas pH qhia tias tsis muaj kev hloov pauv
Kev tshuaj ntsuam xyuas thiab kev soj ntsuam yuam kev: Yog tias tsis muaj qhov hloov pauv hauv qhov ntsuas ntsuas, nws yuav yog vim lub tshuab xa hluav taws xob tsis ua haujlwm lossis kev sojntsuam sib kis, uas tiv thaiv kev ntsuas ib txwm. Lub teeb liab transmitter yog siv los simulate qhov kev sojntsuam thiab xa cov teeb liab mus rau lub transmitter. Lub transmitter tso zis hloov, tab sis lub transmitter yog ib txwm. Tom qab disassembled thiab tshuaj xyuas pH soj ntsuam soj ntsuam, pom muaj kab mob sib kis ntawm qhov kev sojntsuam.


Kev daws teeb meem: Tom qab ntxuav qhov kev sojntsuam, lub ntsuas ntsuas vaj huam sib luag rov qab mus rau qhov qub.


3. Lub ntsuas pH qhia tau tias muaj kev hloov pauv loj heev
Kev tshuaj xyuas thiab kev soj ntsuam tsis raug: Qhov ua rau muaj kev hloov pauv loj hauv kev nyeem ntawv suav nrog kev hloov pauv ntawm lub zog xa hluav taws xob, kev ua haujlwm tsis zoo, lossis teeb meem nrog kev sib txuas ntawm lub transmitter thiab ntsuas ntsuas. Lub teeb liab transmitter yog siv los simulate qhov kev sojntsuam xa cov teeb liab mus rau lub transmitter, thiab cov transmitter tso zis hloov ib txwm. Tsis muaj teeb meem nrog lub transmitter, thiab lub hwj chim mov kuj tseem ruaj khov. Qhov teeb meem yog nyob rau hauv kev twb kev txuas cable los yog kab kev twb kev txuas.


Kev tuav tsis raug: Tom qab hloov cov kab txuas txuas, cov cuab yeej ntsuas tau zoo li qub.


4. Lub iav pH electrode thiab Circuit Court ntawm cov pH meter muaj nyob tsis zoo, tab sis tom qab muab tso rau hauv kev ua haujlwm, qhov kev ntsuas qhov yuam kev yog qhov loj, thiab txawm tias tsis tuaj yeem ua haujlwm li qub.
Kev soj ntsuam yuam kev: Kev tsom xam yog tshwm sim los ntawm ob qho laj thawj. Ua ntej, lub rwb thaiv tsev impedance ntawm shielded cable, hlws ris lub thawv los yog davhlau ya nyob twg ntawm pH electrode thiab lub siab kuj converter yog txo. Vim lub siab sab hauv ntawm iav pH electrodes, rwb thaiv tsev degradation tuaj yeem tshwm sim yog tias cov terminals ntawm pH electrode thiab lub converter ua damp, lub shielded cable yuav moldy, cov tshuaj poov tshuaj chloride siv rau cov ntaub ntawv pH electrode contaminates lub davhlau ya nyob twg lub thawv los yog seeps. mus rau hauv cov ntaub thaiv kab hlau, roj lossis dej phwj uas tseem tshuav ntawm cov terminals thaum lub sijhawm tu, lossis yog tias lub thawv tsis raug kaw thiab cov plua plav hauv nws. Qhov thib ob, muaj cov cuab yeej siv tshuab thiab hluav taws xob loj nyob ze ntawm qhov chaw teeb tsa ntawm lub ntsuas ntsuas, thiab ntau dhau tam sim no cuam tshuam nrog cov ntsuas ntsuas.


Kev daws teeb meem: Siv cov kev ntsuas tsim nyog.


5. Thaum ntsuas pH ntsuas, qhov ntsuas tsis raug
Kev tshuaj ntsuam xyuas thiab kev soj ntsuam yuam kev: Thaum kuaj xyuas, nws pom tau tias cov khoom nruab nrab uas kuaj tau muaj cov roj paug. Qhov no yog vim hais tias thaum nruab nrab muaj ib tug loj npaum li cas ntawm cov roj impurities, nws muaj peev xwm kis tus pH electrode.

 

3 ph meter

Xa kev nug